Correlation Analysis of Human Upper Limb Parameters and Oscillometric Signal in Blood Pressure Measurement
Blood pressure measurement,Oscillometric signal,Upper limb parameters,Vessel-related information,Phantom,FEM modeling,Parametric study
Special Sessions > Human Vibration and Health Sensor Applications
Draft Paper Accepted
Bomi Lee / Korea Advanced Institute of Science and Technology (KAIST)
Yong-Hwa Park / Korea Advanced Institute of Science and Technology (KAIST)

Blood pressure (BP) is one of the most frequently measured bio-signals, and the automatic BP monitor that measures the BP is usually based on the oscillometric method where the pulse along the blood vessel is recorded as a pressure oscillation inside the air cuff. The correlation between oscillometric signal and the upper limb’s bio-mechanical parameters such as radius, thickness, and elasticity of brachial artery, and elasticity of soft tissue is studied to assess the possibility of extracting health condition of humans other than BP during BP measurement. The parametric study is conducted with a simple finite element method (FEM) model containing bone, soft tissue, and the artery to identify the bio-mechanical parameters that affect the amplitude of the oscillometric signal. The result shows that the inner radius of the brachial artery is the most influential parameter on the oscillometric signal. The correlation between oscillometric signal and the inner radius of the brachial artery is also verified by the experiment with the cardiovascular dynamic simulator fabricated by this group [1,2]. The blood pressure waveform generated by the simulator produces the pulse and this is fed into the upper limb phantom to mimic the blood pressure measuring situation. Using series of phantoms with the various artery inner radius, the oscillometric signals are acquired with a commercial automatic BP monitor to evaluate the correlation between the inner radius of the brachial artery and the oscillometric signal. The amplitude variation of the oscillometric signal is also observed from the actual human arm when the vessel vasodilation is induced. The correlation between oscillometric signal and the parameters of the brachial artery shows the possibility of extracting the vessel-related information from the oscillometric signal during BP measurement.

Countdown

  • 00

    Days

  • 00

    Hours

  • 00

    Minutes

  • 00

    Seconds

Important Dates

Abstract Submission Deadline:

 31st March 2021 15th April 2021

Extended Deadline: 1st Aug. 2022

 

Abstract Acceptance:

30th April  2021 Rollover

 

Full Paper Submission Deadline:

30th June 2021  14th July 2021

Extended Deadline: 15th Aug. 2022 

 

Notification of Acceptance:

15th August 2021 1st Sept. 2021

1st Sept. 2022

Contact Us

  Tel: 86-0532-6897 5191 (Ms Yuan)

  Mob: 184 5327 6561
  E-mailsecretariat@apvc2021.org
               organizer@apvc2021.org

Visitors